Displaying report 1-1 of 1.
Reports until 10:05, Thursday 28 February 2019
H1 SQZ (SQZ)
nutsinee.kijbunchoo@LIGO.ORG - posted 10:05, Thursday 28 February 2019 - last comment - 15:27, Friday 01 March 2019(47184)
Two out of four acoustic peaks gone

The peaks at 206Hz and 234Hz are gone. I guess the PBS did something good. The other two lines are still persistent. One at 177.7Hz got worse after I closed and reopened the beam diverter after a long lock (orange trace for ASQZ, cyan for SQZ). Compare to purple trace that I took right when I got in the control room. The line at 262Hz got worse after the beam diverter closed/opened but got better after sometime passed (dark purple). Maybe this suggests something mechanical? The next thing to do is to check the coherence with accelerometers.

 

Note that I kept the phase the same as last sqz measurement and I didn't optimized the nlg before I took this measurement (temperature setting was the same as last time).

 

Images attached to this report
Comments related to this report
chris.whittle@LIGO.ORG - 15:27, Friday 01 March 2019 (47229)SQZ

You could try taking another spectrum with the noise eater turned off. At LLO we found the noise eater was adding a peak at ~200Hz to our CLF (aLOG 43780).

Displaying report 1-1 of 1.