I injected bandlimited noise into H1:ASC-DHARD_P_EXC starting at about 23:55:00 UTC 8 Mar 2019, and increasing the size of the excitation by 3x at about 23:58:00 UTC, and then stopped the excitation around 00:01:00 UTC (9 Mar 2019). Hopefully this is enough data to see if the DHARD coupling to DARM is more stationary.
EDITED to correct the time zones.
At a first look, DHARD_P coupling appears to be much more stationary. Comparing to 47274:
I also ran the non-stationary noise subtraction code, and confirmed that most of the noise coupling is stationary (see 4th figure). The largest non stationary contribution comes from modulation due to PRC1_P