NikoL, RickS
In response to comments that the 14 Hz harmonic of the Pcal injection at 7 Hz is larger (relatively) than it was in the past (it is still about -100 dBc), we adjusted the Optical Follower Servo (OFS) offset from the control room as follows:
When we started, the loop offset was at 3.0, we had dropped it from 3.75 earlier in the day. We left it at 2.95. This is a ~20% decrease that might indicate that the laser output power is dropping, or that the AOM diffraction efficiency has decreased, or a combination of both.