Reports until 13:06, Tuesday 12 March 2019
H1 ISC
richard.mccarthy@LIGO.ORG - posted 13:06, Tuesday 12 March 2019 (47460)
EX safety system

While trouble shooting the glitches on the PCAL temperature sensor a two connectors at EX were inadvertently swapped.  This caused a problem when the pcal lid was opened and the ALS table door was opened. The system thought it was in an unsafe condition and tripped off the lasers.  I have fixed the wiring issue that existed at end X only.  Y-end was not a problem.