Reports until 17:40, Tuesday 20 November 2012
H1 CDS
david.barker@LIGO.ORG - posted 17:40, Tuesday 20 November 2012 (4750)
BSC1 front end code started

WP#3571

The SUS and ISI models to support BSC1 install work were started today. These were non-standard installs due to the following constraints:

Both of these limitations are temporary. When we upgrade to RCG2.6 the IPC limit will be raised. Tomorrow I will receive more memory for the H1DAQ (most probably wont install till next Tuesday).

I created IOP models for the SUS and SEI BSC1 frontends (h1susb123 and h1seib1). These were copies of the H2 equivalents minus the Dolphin IPC.

I copied H1SUSETMY model to become H1SUSITMY. The appropriate DCUID/computer changes were made and all Dolphin IPC was removed.

I copied the H1ISIETMY model to become H1ISIITMY. Again DCUID was changed and Dolphin IPC was removed. We had some confusion about the ADC distribution between HEPI and ISI. The models we have been using give ISI the first three ADCs and HEPI the fourth. Drawings indicate the opposite, HEPI getting the first ADC and ISI the last three. We will work on resolving this with team SEI.

As stated earlier, no H1DAQ changes were made. None of the BSC1 channels are being recorded by the DAQ.