H1 ISC (CAL, DetChar)
anamaria.effler@LIGO.ORG - posted 16:32, Sunday 17 March 2019 (47600)
PUM and UIM calibration tracking is not quite right

DanB, Sheila, Anamaria

As reported before, eg here, the calibration is pretty good down to some 30 Hz according to pcal broadband injection. kappa_pum and kappa_uim are reporting very different numbers from the nominal gains of 1 in the analog suspension calibration filters (~0.8) so we thought maybe these things are connected. We implemented them in the calibration filters of ANALOG ETMX L2 and L1, but after a pcal sweep, that made the low frequency calibration much worse; and the 35 Hz - ish pcal line visibly did not match anymore. We backtracked these changes.

Kappa_tst is 1 and the filter is 1, so at least that matches well.