Displaying report 1-1 of 1.
Reports until 17:36, Wednesday 20 March 2019
H1 CAL
sheila.dwyer@LIGO.ORG - posted 17:36, Wednesday 20 March 2019 (47723)
calibration measurement lockloss

We lost lock at around 0:30 UTC because of the lowest frequency points in the DARM OLG measurement.  This is actually the second time we've lost lock in the last ~week or so running this measurement, so I've reduce the amplitudes of the points below 7Hz. 

Displaying report 1-1 of 1.