Displaying report 1-1 of 1.
Reports until 15:43, Monday 25 March 2019
H1 TCS (ISC)
jenne.driggers@LIGO.ORG - posted 15:43, Monday 25 March 2019 (47851)
Ring heater test from Saturday night - common higher is not better

I had a brief look at some data from a short ring heater test that was done Saturday night.  It's not alogged when the change was made, but Ed alogged when he did the step back (alog 47797).  It looks like both ITM ring heaters were increased by 0.2 W total (0.1 W upper, 0.1 W lower) for one hour, and then lowered back to their nominal values. 

Attached is a screenshot showing the step times, some of the circulating powers and power recycling gain, as well as spectra taken at 2 times. The spectra are from the times of the 2 vertical cursors on the time series plots. The step up in ring heater power was done about an hour into the lock, so there are still some thermalization effects going on. But the step down an hour later helps make things more clear.  In the spectra, red is the hotter ring heater state, and blue is the nominal ring heater state.  Recall for the spectra that we'd like all of the peaks to be smaller, indicating a lower coupling of that noise to DARM.  For POP90 we'd like it low, and for PRG and POP18 we'd like them high.

So, overall, I conclude that the nominal common ITM ring heater values are better than the hotter ring heater values.  This is opposite of what one might expect with the increase in total circulating power in the IFO, if we assume that we had fine tuned the TCS for our previous 30W settings. But, we hadn't actually done the fine tuning for TCS at 30W. 

 

Images attached to this report
Displaying report 1-1 of 1.