H1 SQZ (SQZ)
nutsinee.kijbunchoo@LIGO.ORG - posted 22:28, Monday 25 March 2019 - last comment - 15:30, Tuesday 26 March 2019(47865)
CLF to DARM injection

Craig, Georgia, Nutsinee

It seems like we have non-linear coupling from CLF to DARM. The coherence plot wasn't very obvious but we clearly saw out injection showed up everywhere in DARM as we swept. Lockloss before we could inject more squeezing measurement (it's not the squeezer!).

Images attached to this report
Comments related to this report
lee.mcculler@LIGO.ORG - 11:48, Tuesday 26 March 2019 (47885)

Hmm. How hard were you driving? I suppose it's either fringe-wrapping or nonlinear upconversion of the backscatter. At LLO, driving the ZM's also shows such nonlinear conversion into DARM LLO43807

The relevant ZM1 drive plot in DARM.(DARM calibration not setup here, but you can clearly see the 2F and even 3F harmonics for a small 1F drive)

 

 

daniel.sigg@LIGO.ORG - 15:30, Tuesday 26 March 2019 (47900)

I assume we are modulating the back scattered light, so the phase wanders around. We will have to repeat this with a power spectrum peak height ratio measurement.