georgia.mansell@LIGO.ORG - posted 03:30, Tuesday 26 March 2019 - last comment - 19:26, Tuesday 26 March 2019(47869)
Ring heater differential step started at 10:29 UTC

Following on from Jenne's common ring heater change (lowering power), we are taking a differential ring heater step tonight. We will reduce ITMY to the same power as Jenne did, and take a smaller step on ITMX. If there are problems with ASC while reacquiring lock, these changes might need to be reverted.

  ITMX RH Filter input [W] ITMY RH filter input [W]
Nominal 0.5 1.4
Common step today 0 0.9
Differential step tonight 0.25 0.9

( - I'm quoting filter input requested power here, there's an offset from the actual ring heater set power I'm not sure about.)

 

Craig has coded up a live monitor of the frequency and intensity noise lines which I'm running on ZOTWS22. The realtime monitors are found here:

    /ligo/home/craig.cahillane/monitors/AWG_Line_Monitors.py

To run them you might need to use craig's environment, by running:

    source /opt/rtcds/userapps/release/cds/h1/scripts/setup_anaconda

    source activate cragenv

 

(edit: original title to alog quoted wrong start time timezone for measurement)

Comments related to this report
jeffrey.bartlett@LIGO.ORG - 07:08, Tuesday 26 March 2019 (47871)
   After ASC lockloss reset the Ring Heaters back to original values. RH Filter Input for ITM-X is now 0.5w, RH Filter Input for ITM-Y is now 1.4w. 
georgia.mansell@LIGO.ORG - 19:26, Tuesday 26 March 2019 (47914)

Just to clarify: the nominal ring heater values I quoted in this alog are 0.1 W higher than the actual nominal ring heater vales: I got confused by the slow response of the filtered output.

Before yesterday, ITMX_RH_INVERSE_FILTER_IN was 0.4 W, and ITMY_RH_INVERSE_FILTER_IN was 1.3 W.