During the quiet lunch hour Mark ran 3 of the 6 DOF TFs on the ITMy main chain. The TFs indicated the ITMy looked healthy with resonance peaks in appropriate places. At 4pm, Vincent reported that he could not run ISI measurements because he ITMy watchdog was tripped. A quick investigation by Mark, Vicent, Hugh and I on DV indicates that indeed the ITMy seems rung up in Transverse (a motion only driven by the single side BOSEM). As well, while we can enable the damping loops and watchdaogs such that signals wiggle out the far right of the ITMy MEDM, we aren't convinced damping is actually happening on the ITMy as seen by BOSEM channels on DV, and a quick noisy TF of longitudinal motion. No mechanical work was undertaken on the ITMy today. Possible culprits:
SEI work up top in the morning and afternoon could have jarred the ITMy into an EQ stop - will check in the morning, but doubtful as the BOSEM (sensors) postions look unchanged from the morning.
IOP work to remove BS which was hindering measurements, including frontend and model reboots/restarts to everything related to the IOP - Maybe we need a fresh reboot/restart? Will press on with troubleshooting tomorrow.