So far OPO crystal temp of 33.337C is a good place for .945 mW requested power at OPO refl when locked. The measured NLG is 2.145.
----------------------------------------------
Our laser is running multimode again. The controller has been adjusted to move away from the multimode region on March 12 (alog47481). Last time our laser tripped and this time I opened the table to clean up and take some close out photos. I wonder if the laser is some how very sensitive to temperature. For now I'm just going ot leave it and see if it will stop multimoding after the temperature has settled down. With our intensity stabilization servo on I don't think we are suffering noise wise.