Intermittent hardware errors on h1ecaty1

We are seeing intermittent Beckhoff errors at EY, Daniel is investigating.

Images attached to this report
Comments related to this report
daniel.sigg@LIGO.ORG - 14:46, Wednesday 03 April 2019 (48210)

We are temporarily loosing communication with terminals M11 through M17 of  the end station 2 chassis. M11 is a a EL9410 power supply terminal. This could indicate that this module is failing. It could also indicate that the prior terminal, EL4134, has a problem communicating with the next terminal.

Problem started 4/3/2019 19:39 UTC.

Images attached to this comment
daniel.sigg@LIGO.ORG - 14:58, Wednesday 03 April 2019 (48211)

Fil and Marc swapped the power supply terminal.

Fault: https://services.ligo-la.caltech.edu/FRS/show_bug.cgi?id=12659