david.barker@LIGO.ORG - posted 08:32, Friday 05 April 2019 - last comment - 11:51, Sunday 07 April 2019(48260)
h1calcs crontab to diag_reset not working
My python script to diag_reset h1calcs when it occasionally runs long is not working as a cronjob. For now I'm running it in a tmux session on h1fescript0 (looping every minute) until I can get the cronjob functional.
Comments related to this report
david.barker@LIGO.ORG - 11:01, Friday 05 April 2019 (48264)
attached image of h1calcs cpu_meter_max shows an automatic reset event at 17:15 UTC (10:15 PDT) Saturday morning. Within the same minute the cpu went to 63uS (the TIM bit was set on STATE_WORD) and then DIAG_MAIN was reset.
Running again as a cronjob once I got the environment sorted out.
attached image of h1calcs cpu_meter_max shows an automatic reset event at 17:15 UTC (10:15 PDT) Saturday morning. Within the same minute the cpu went to 63uS (the TIM bit was set on STATE_WORD) and then DIAG_MAIN was reset.