Busy morning. Lock loss just after "LSC Servo Board Split Voltage High" Message. Had GRB alert while relocking the IFO Rode through a mag5.7 EQ in Chile - SIESMON said to select EARTHQUAKE state. Did not change state and IFO continued to locking at NLN with no apparent issues. Dropped out of Observing due to OPO SDF Diffs. ZM1 UL, LL and ZM2 UL, LL, UR, LR saturating. Looked like Squeezer was still locked. Called Keita. He cleared the Servo Filters on the SQZ_ASC Overview, relocked the SQZ_SHG, and requested SQUEEZING from the SQZ_MANAGER. SDF Diffs cleared, and the IFO is back into Observing. Right now the IFO is locked in NLN with a range of 110.6Mpc. Winds are a no more than a Light Breeze. Am tracking 5 increasing Violin modes (ETM-Y mode2, mode4, mode8 & ETM-X mode6 & ITM-Y mode12). All are under 2.5 and growing slowly if at all. Will continue to monitor.
About the squeezer issues:
I haven't loaded either of these changes, since we are in observing right now, but if there is any reason to come out of observing tonight I think that it should be OK to load both of them.