daniel.vander-hyde@LIGO.ORG - posted 20:37, Tuesday 09 April 2019 - last comment - 11:02, Tuesday 16 April 2019(48349)
Second point absorber mask installed (first one removed)

Georgia, Danny

With different ITM spot positions we have changed how the point absorbers contribute to the overall optical path distortion. This makes the first CO2Y mask design (Mask 1) less helpful, Aidan designed a second iteration of the CO2Y mask (Mask 2) in order to better accommodate for the new optical path distortion. We installed this new mask today into the annular mask beam flipper.

We believe the mask should be installed in the flipper as indicated in the first figure

Some notes on the installation of Mask 2:

Images attached to this report
Non-image files attached to this report
Comments related to this report
daniel.vander-hyde@LIGO.ORG - 20:49, Tuesday 09 April 2019 (48367)
Images attached to this comment
daniel.vander-hyde@LIGO.ORG - 11:02, Tuesday 16 April 2019 (48533)TCS

Made some fine adjustments to the installed mask (Mask 2). A FLIR image is attached. Also attached are Hartmann images comparing the centering of the optical path distortion from the point absorbers (left) and the optical path distortion of CO2Y with the mask in its current alignment. It is important to note that the optical path distortion shown in the HWS image to the right is caused by a short duration 2.4W CO2 step up and not fully representative of the distortion you expect to see at thermal equilibrium. 