Marc Nutsinee Daniel
All common mode boards in the squeezer rack (CLF S/N S1700344; OPO S/N S1700345; SHG S/N S1700346; LO S/N S1700348) as well as the 2 spares (S/N S1700347 & S/N S1700349) were modified according to E1900103 (high pass filtering of DAQ channels to avoid slew rate limitations in the AA chassis).
Installed the AM modulated AOM driver chassis D1900045 (S/N S1900202) and removed the temporary ifr RF synthesizer and Mini-Circuits RF amplifier.
Updated the model to add a dedicated integrator at the output of the green pump power stabilization. This prevents the filter output from increasing to very large numbers, when the actuator is out of range. The servo can now be turned on and off with a trigger from the OPO transmitted power.
When we tried to relock the squeezer, we had to fix several problems with broken equipment!? Cause unknown.
We replaced the PZT driver with the spare (S/N S1700172) which fixed the offset problem for the SHG.
We checked the CLF drive at the output of the AI board and found the same problem: both legs of the differential signal are pegged at -13V. This eliminates the cable as a problem.
Attached transfer functions of LO, OPO, and SHG board. CLF excitation didn't work.