[JeffK, Jenne]
Jeff noticed that the actuator kappas looked funny, and determined that we had another timing slip between CAL CS and SUSETMX for the synced oscillators, like in alog 48129. I zero-ed the frequency of the oscillators and reset them to their nominal values, and that fixed the situation again.
Since I had forgotten to save a template last time, the template that generated the attached screenshot is at /ligo/svncommon/CalSVN/aligocalibration/trunk/Runs/O3/H1/Measurements/CALCS_FE/2019-04-10_H1CALCS_SUSOscillatiorSlip_Diagnotic.xml
Longer term, since Jeff's proposed solution from the FRS (12335) was successfully implemented at LLO (LLO 45007) yesterday, we'll make the same change here at LHO next Tuesday, and will hopefully never have to worry about these timing slips again.