Getting rid of CAL OSC timing slips; send oscillator signal from ends to corner

I have begun some prep work for tomorrow's maintanance day installation of updated CALCS and SUS quad models under work permit 8166.  This will address FRS 12335, as laid out in ECR E1900107, to eliminate the problem of timing slips on the supposed-to-be-synced oscillators causing problems with the time dependent calibration kappa calculations.  JoeB successfully completed this at LLO last Tuesday (LLO alog 45007), so here we are just pulling his changes.

So far, I have:

To do tomorrow during maintanance:

Comments related to this report
jenne.driggers@LIGO.ORG - 17:53, Monday 15 April 2019 (48514)

For the phase delays to account for the 1 cycle delay with the new oscillator, the appropriate amount of delay in degrees depends on the line frequency. Here I list the delays for both our current calibration line frequencies, as well as for the proposed frequencies (see alog 48512).

The delays will go in channels of the type H1:CAL-CS_TDEP_SUS_LINE{1,2,3}_SUS_DEMOD_PHASE, and are calculated from angle(exp(-2*pi*i*Freq/16384))*180/pi.

15.1 Hz = -0.3318 deg

16.7 Hz = -0.3669 deg

35.9 Hz = -0.7888 deg

15.6 Hz = -0.3428 deg

16.4 Hz = -0.3604 deg

17.1 Hz = -0.3757 deg.  (EDIT: this is the freq for the pcal line, which needs its delay updated)

17.6 Hz = -0.3867 deg. (EDIT: this is the freq for the sus line, which we need to put the delay in)

Also, since we are using ETMX for all of our DARM actuation, we will set channels of the type H1:CAL-CS_TDEP_SUS_L{1, 2, 3}_LINE_END_SW equal to 1. 0.  EDITED - 0 is for EX (which we want), 1 is for EY (we do not want).

jenne.driggers@LIGO.ORG - 09:42, Tuesday 16 April 2019 (48529)

Since we are rebooting these models (and I remembered that the temporary channels were in there), I have removed the following channels from the quad suspensions:

H1:SUS-ETMX_L2_TEMP_P_1            1   0     0   4     16   0
H1:SUS-ETMX_L2_TEMP_P_2            1   0     0   4     16   0
H1:SUS-ETMX_L2_TEMP_P_3            1   0     0   4     16   0
H1:SUS-ETMX_L2_TEMP_TP_2           1   0 10814   4  16384   0
H1:SUS-ETMX_L2_TEMP_TP_3           1   0 10813   4  16384   0
H1:SUS-ETMY_L2_TEMP_P_1            1   0     0   4     16   0
H1:SUS-ETMY_L2_TEMP_P_2            1   0     0   4     16   0
H1:SUS-ETMY_L2_TEMP_P_3            1   0     0   4     16   0
H1:SUS-ETMY_L2_TEMP_TP_2           1   0 10814   4  16384   0
H1:SUS-ETMY_L2_TEMP_TP_3           1   0 10813   4  16384   0
H1:SUS-ITMX_L2_TEMP_P_1            1   0     0   4     16   0
H1:SUS-ITMX_L2_TEMP_P_2            1   0     0   4     16   0
H1:SUS-ITMX_L2_TEMP_P_3            1   0     0   4     16   0
H1:SUS-ITMX_L2_TEMP_TP_2           1   0 10798   4  16384   0
H1:SUS-ITMX_L2_TEMP_TP_3           1   0 10797   4  16384   0
H1:SUS-ITMY_L2_TEMP_P_1            1   0     0   4     16   0
H1:SUS-ITMY_L2_TEMP_P_2            1   0     0   4     16   0
H1:SUS-ITMY_L2_TEMP_P_3            1   0     0   4     16   0
H1:SUS-ITMY_L2_TEMP_TP_2           1   0 10814   4  16384   0
H1:SUS-ITMY_L2_TEMP_TP_3           1   0 10813   4  16384   0

None of these were being saved to frames.  These had been put in, in order to diagnose some problems with the IFO lockloss trigger, and should have been removed months ago.  However, since we don't ever want to reboot the suspension models unnecessarily and these channels weren't taking up frame space, we had left them in.

I have re-compiled and done make-installs for all 4 quad models.  I have also done a make-install for h1calcs (it hasn't changed since I compiled yesterday afternoon).

jenne.driggers@LIGO.ORG - 11:34, Tuesday 16 April 2019 (48531)
  • SDF checked - no diffs for CALCS, accepted in safe.snap the slider values and A2L gains for all 4 quads.
  • Pulled medm changes for CALCS
    • CAL_CS_TDEP_OVERVIEW.adl,  CAL_CS_TDEP_SUS_LINE.adl,   Updated to revision 19395.
  • JeffB took quad suspensions to SAFE, but we forgot to put the ISI platforms to DAMPED so those platforms tripped.  Dave restarted all 5 models.
  • Demod phases for single clock cycle installed and accepted in SDF safe and observe.
    • JeffK reminded me that the Pcal Line1 demod needed its delay value updated to match the new line frequency.  This wasn't mentioned in JoeB's LLO alog, so I had forgotten it.  Thanks Jeff!
  • JeffK is changing the line frequencies, and updating bandpass filters to match.