Sheila, Danny
Today we took about 5 hours out of observing in order to do a series of OMC scans. The measurements were heavily inspired by what Dan Hoak did in alog 22175.
What I wanted to see from these scans was a direct effect of the high spatial frequency optical path distortion on the carrier and 9 MHz and 45MHz sidebands. My attempt to accomplish this required the following:
Another separate but relevant measurement was to measure the field content at the AS port with a 35W thermal state. Because we did not want to damage the PD by scanning at 35W we let the interferometer thermalize at 35W and dropped down to 10W to take scans. It is important to note that it was not trivial to power up to 35W on DC readout and transition back to RF hence the large amount of time out of observing. Because of time constraints, the thermalization criteria had to be an hour and a half at 35W rather than 2+ hours which should be good enough for this measurement.
Results from the scans are pending and will be posted in a comment on this log entry.