Ops Owl Mid-Shift Summary
   Trying to align and lock the IFO. Wind is mostly in the low to mid teens, with gusts 20mph Plus. Although coming down, the microseism is a bit elevated, (Y 0.9um/s, X 0.09um/s, Z 0.09um/s).

   Trying to run initial alignment.
 
   Can align and lock the ALS X & Y green easily, but it will not Offload. 
MSG: "GREEN_WFS_OFFLOAD.enter" and no further. The ALS_X_WFS_DOF_2 and ALS_Y_WFS_DOF_2, both in both pitch and yaw appear not to be settling down.  

   After waiting 30 plus minutes for the Offload to complete, skipped to Align Input Laser Pointing to Xarm. This step completed without trouble. Could not get past PRX_LOCKED. Tried tuning PRM until I had what looked like a good AS_AIR spot. But still could not get to PRC_ALIGN. 

   Restarted the initial alignment and still cannot get ALS to unload. Am continuing trying to get through initial alignment.          