posted 06:49, Saturday 04 May 2019 - (48975)
Symptoms:
- Do not see a spot for ALSy on the video
- Error signal for CAM ITM PIT is above 0.2 (it is at -1.2). Trouble shooting document said I should Clear History, but that does not appear to work any more (the file for it is no longer in the folder....so maybe this is an old solution?)
- Clear History file was: /opt/rtcds/userapps/release/asc/h1/scripts/softCH_arm.py (but softCH_arm.py is no longer in that folder)
- Continue investigating
I forgot about Jenne's note about changing AS_A_36 phase. So I have done that, but it didn't address the ALSy issue.
Trending oplevs for ETMy & ITMy does show some big alignment changes. I still wonder if this is related to a WFS issue. Will steer optics back to see what this does.
NOTE: I have not seen this issue before with a Y-arm misalignment like this!
Ed's here to save the day, but I'm trying to remember what I saw & did with ALSy.
Definitely saw ITMy & ETMy both shift their alignment! (see attached to see ITMy/ETMy pit/yaw shifts & various things I did to bring back).
Back Story:
Wanted to also mention that I clicked on the "Clear History" for each of the individual Filter Banks on the ALS Y Overview (which I assume the non-functioning yellow Clear History button would do on this medm window). When I did this I did not notice any change to the CAM ITM PIT Error signal. These filter banks had their inputs OFF and their outputs were zero. Just wanted to note that I did click their Clear Histories though.