Reports until 16:30, Monday 06 May 2019
H1 CDS
jonathan.hanks@LIGO.ORG - posted 16:30, Monday 06 May 2019 (49045)
Enabling auto restart of the CW injection on failure with ramping
Dave, Jonathan,

During earthquake down time we reworked systemd config controlling the CW injection.  We make sure that startup/shutdown of the injection enforces a 10s ramp time.  The system will try to auto start up to 4 time in 1/2 hour, with 5 minutes between attempts.

We changed the startup script to:

source /ligo/cdscfg/stdenv.sh
caput H1:CAL-INJ_CW_TRAMP 10.0
caput H1:CAL-INJ_CW_GAIN 0.0
sleep 11
/home/hinj/Details/bin/x_start_psinject monit start
caput H1:CAL-INJ_CW_GAIN 1.0


and the stop script to:

source /ligo/cdscfg/stdenv.sh
caput H1:CAL-INJ_CW_TRAMP 10.0
caput H1:CAL-INJ_CW_GAIN 0.0
sleep 11
/home/hinj/Details/bin/x_stop_psinject monit stop


As a side effect of how systemd monitors it, when the process crashes, the stop script gets run which ensure that on a failure the gain is forced down to 0 with a ramp. So whatever was left in the awg buffers gets ramped down.