Installed a 50/50 BS in the camera path, to reduce the power on the camera at 35W IMC IN power. The camera was seeing 30uW, and is now seeing ~17uW.
With IMC IN at 10W, the incident power on the 50/50 is 4.6uW, and the transmitted power is 5uW.
- Cheryl, Jason