Reports until 00:23, Wednesday 08 May 2019
H1 CDS
david.barker@LIGO.ORG - posted 00:23, Wednesday 08 May 2019 - last comment - 00:46, Wednesday 08 May 2019(49091)
Corner station Dolphin failure

Looks like all front ends on the corner station dolphin network are having issues, I'm starting the recovery.

Images attached to this report
Comments related to this report
david.barker@LIGO.ORG - 00:30, Wednesday 08 May 2019 (49092)

I've captured dmesg and dolphin logs for later analysis, I'm now restarting all the models in the order SUS, SEI, ISC.

david.barker@LIGO.ORG - 00:46, Wednesday 08 May 2019 (49095)

All models restarted with no issues. 

The CW psinject had stopped because h1calinj was restarted, and it sucessfully restarted itself now that it is under systemd control.

I cleared all the IPC errors and DAQ-CRC counters. Attached CDS Overview shows that the only red is due to the suspension models still have watchdogs enabled.

I'm handing the system over to Corey for IFO recovery.

Images attached to this comment