TITLE: 05/08 Owl Shift: 07:00-15:00 UTC (00:00-08:00 PST), all times posted in UTC
STATE of H1: Corrective Maintenance
INCOMING OPERATOR: Jeff
SHIFT SUMMARY:

This shift went super fast....thanks to a dolphin crash!  Ha.  But it was actually kind of fun to go through all the problems; definitely a learning experience!  Sadly, because I am slow I am only handing off an H1 with DRMI locked.

Back Story:  Just as Niko was about to leave after his handoff from me, had a major dolphin crash in the LVEA (Dave will have a more informed entry on the specifics of the crash...he did mention we hadn't had one of these since January where we had been suffering from these fairly regularly before then).

I tried to post specifics & timing of events in other alogs for what I found and did during my shift for this recovery.  In broad strokes, here were my method to the madness and areas I focused on in order through the night:

  1. Proceed as far as I could with locking as usual, until I ran into an issue.  (I was stopped early due to IR cavity for X!)
  2. Alog search for similar issue (Basically looked at last instances for recovery from Dolphin recoveries (thanks, Sheila, Dan, & JimW for the two entries I found).
  3. Dark Offsets!  (I should have realized this was an issue early on (TR_X [blue on striptool] was up at 0.3 instead of 0!)
  4. Check pit/yaw sliders for SUS (If I did this sooner (via Time Machine?) I could have saved some time scratching my head about why certain cavities were not locking.))
  5. ISI Config!  (If I didn't read about this in the Dan/Jim alog, I would have been dead in the water.)