I just noticed that the h1calcs model is using its safe.snap file as its SDF reference and not OBSERVE.snap.
Looking at trends and logs, it appears that the following has happened (all times local):
Fri 05apr2019 16:01: switch_SDF_source_files.py was modified to add calcs to the exclude list, so guardian is not controlling its sdf reference
Tue 16apr2019 10:52: h1calcs model restarted, OBSERVE.snap reference set.
Wed 08may2019 00:40: h1calcs model restarted (Dolphin crash), safe.snap reference set (this is the current state).
There are differences between safe.snap and OBSERVE.snap (see attached text file).
Keita is in contact with the calibration group on this issue.