Reports until 04:32, Thursday 06 June 2019
H1 General (OpsInfo)
corey.gray@LIGO.ORG - posted 04:32, Thursday 06 June 2019 - last comment - 05:46, Friday 07 June 2019(49693)
H1 Lockloss Recovery From 37W Lockloss at 8:16 (Operator Play By Play)

Lockloss at 8:16  (lockloss file)

Reverted to 35W per Sheila's alog and noted in my earlier "Back to Observing" alog.  Here is the "Operator Play By Play" for locking. 

NOTE:  This is after two new states:

  1. Higher Power (37W) lockloss  
  2. New Operational changes to ALS & DRMI from Tuesday.  (Sheila requested operators to give a play by play for how handling these new changes affects locking.)

Play By Play Notes:

Additional NOTE:  I wasn't sure whether to mark downtime as "locking" or "commissioning".  I did a little bit of both, but mostly "commissioning", because I had to undo some high power stuff and then had to deal with new procedure for ISC_LOCK from Tues work.

10:04 Back to OBSERVING

Images attached to this report
Comments related to this report
thomas.shaffer@LIGO.ORG - 09:56, Thursday 06 June 2019 (49707)

In regards to the ALS_X/YARM in UNLOCKED, on the main guardian screen for a node, there is a Request area below State and Target. This area has 2 fields, an EPICS readout and a state selector. The state that the state selector reads is not an accurate way to judge what state we are in, hence the EPICS readout right above it. There are only so many states that can fit on the state selector, so if a state is requested from a manager that is not in the selector, it will not be able to show it there. Recent changes have added new states to this node, most likely pushing Unlocked off of the state selector.

Hopefully this clears it up a bit.

corey.gray@LIGO.ORG - 05:46, Friday 07 June 2019 (49732)

Thanks for the explanation, TJ!  

I really wanted to note/document what I saw because it was different from what I had been used to & so it had me scratching my head a little.  But I do remember there were changes made on Tuesday.  Anyway, I've had experience with the new way it works now and think I have it down.  THANKS!