Checked the dust monitor vacuum pumps and the HEPI pump stations at both ends and the corner station.
Dust Monitor Vacuum Pumps - Made slight adjustments in the vacuum pressure on all three pumps. The temperatures, while all were within specifications, they were a bit high at End-X. The pump housing and exhaust temps were a few degrees above 150f. This is not an issue, but will be if the temps continue to rise. Will watch for a trend developing.
HEPI Pump Stations: All three HEPI pump stations fluid levels had not changed since the last check. I noted no leak or drips that Hugh had not all ready flagged and put absorbent pads down.
The station levels are: End-X - 8 9/16", no change from last check
End-Y - 9 0/16". No change from last check
CS - 6 9/16". Up 1/16" from last check.
Closing FAMIS #12989.