H1 ISC
jenne.driggers@LIGO.ORG - posted 16:45, Monday 10 June 2019 - last comment - 15:12, Wednesday 12 June 2019(49783)
Power up today

Summary: IFO running at 37 W PSL input.  ETM ring heaters increased to avoid PI ring ups (so far, successful). LSC FF measurements done. CAL sensing function measurements done. (Note: New feedforward filters are not installed, I've just taken measurements for the fits).

Comments related to this report
georgia.mansell@LIGO.ORG - 19:43, Monday 10 June 2019 (49796)TCS

I took a quick look at the ITMY HWS, comparing last night's lock (first attachment) and today's lock (second attachment). Last night we had 35W input power, today we have 37W and new spot positions (A2L gains shown in the attachments).

The two point absorbers are clearly visible in the contour and quiver plots. It looks like with the updated spot positions we might have moved further from the main absorber and are sitting between it and the smaller absorber above it. This seems consistent with the ITMY P2L gain being further from zero.

The other Hartmanns are not as clear as ITMY.

Images attached to this comment
jeffrey.kissel@LIGO.ORG - 15:12, Wednesday 12 June 2019 (49859)CAL
Sensing function measurements have been saved and committed to here:
    /ligo/svncommon/CalSVN/aligocalibration/trunk/Runs/O3/H1/Measurements/FullIFOSensingTFs/
        2019-06-10_H1_DARM_OLGTF_LF_SS_5to1100Hz_15min.xml
        2019-06-10_H1_PCAL2DARMTF_LF_SS_5t1100Hz_10min.xml