Reports until 11:18, Saturday 22 December 2012
LHO VE
kyle.ryan@LIGO.ORG - posted 11:18, Saturday 22 December 2012 (4999)
Switched HAM2-HAM3 pumping over to Turbo
1030 hrs. local -> HAM2-HAM3 @ 1.2 x 10-1 torr -> Relatively fast rough-down (scroll pump) due to 2" hose connection to 2.5" Vent/Purge valve as opposed to pumping through deenergized turbo pump (observed ultimate pressure of this particular scroll pump is 3 x 10-3 torr).  

Tumbleweed crew bailing near Y-arm penetration.  

1115 hrs local -> Leaving site now.