Reports until 18:08, Tuesday 18 June 2019
H1 PSL
jeffrey.bartlett@LIGO.ORG - posted 18:08, Tuesday 18 June 2019 (50050)
PSL Weekly (FAMIS #11017)
    Laser Status:
    Front End Power is 31.95W (should be around 30 W)
    70W Output Power is 70.23W
    Front End Watch is GREEN
    70W Watch is GREEN

    PMC:
    It has been locked 21 days, 6 hr 46 minutes (should be days/weeks)
    Reflected power = 10.43Watts
    Transmitted power = 54.28Watts
    PowerSum = 64.71Watts.

    FSS:
    It has been locked for 0 days 2 hr and 57 min (should be days/weeks)
    TPD[V] = 4.935V (min 0.9V)

    ISS:
    The diffracted power is around 2.1%
    Last saturation event was 0 days 2 hours and 57 minutes ago (should be days/weeks)


    Possible Issues: