J. Kissel I gathered the following regular calibration measurements today; in summary -- the full suite. I was also able, during the sensing function portion of the measurements, to grab complementary PCALX sweeps and broadband measurements as well to further facilitate studies of the apparent a-causality of the detuned SRC spring. Total measurement time was 3.1 hours. /ligo/svncommon/CalSVN/aligocalibration/trunk/Runs/O3/H1/Measurements/FullIFOSensingTFs/ 2019-06-26_H1_DARM_OLGTF_LF_SS_5to1100Hz_15min.xml # 1/(1+G) 2019-06-26_H1_PCALX2DARMTF_LF_SS_5t1100Hz_10min.xml # C_x/(1+G) 2019-06-26_H1_PCALY2DARMTF_LF_SS_5t1100Hz_10min.xml # C_y/(1+G) 2019-06-26_H1_PCALX2DARMTF_BB.xml # from 2019-06-27 00:52:26 - 00:54:09 2019-06-26_H1_PCALY2DARMTF_BB.xml # from 2019-06-27 00:49:25 - 00:51:09 2019-06-26_H1_OMCDCPDSUM_to_DARMIN1.xml # TF to roughly convert units from DARMIN1 ct to OMC DCPD current /ligo/svncommon/CalSVN/aligocalibration/trunk/Runs/O3/H1/Measurements/FullIFOActuationTFs/ 2019-06-26_H1SUSETMX_L1_iEXC2DARM_10min.xml # A_L1*C_y / (1+G) 2019-06-26_H1SUSETMX_L1_PCAL2DARM_8min.xml # C_y / (1+G) for L1 2019-06-26_H1SUSETMX_L2_iEXC2DARM_22min.xml # A_L2*C_y / (1+G) 2019-06-26_H1SUSETMX_L2_PCAL2DARM_8min.xml # C_y / (1+G) for L2 2019-06-26_H1SUSETMX_L3_iEXC2DARM_8min.xml # A_L3*C_y / (1+G) 2019-06-26_H1SUSETMX_L3_PCAL2DARM_8min.xml # C_y / (1+G) for L3 Analysis to come.