Displaying report 1-1 of 1.
Reports until 00:17, Wednesday 24 July 2019
H1 General
yannick.lecoeuche@LIGO.ORG - posted 00:17, Wednesday 24 July 2019 - last comment - 02:51, Wednesday 24 July 2019(50764)
Ops OWL Shift Transition

Ops Shift Transition: 07/24/2019, Owl Shift 07:00 – 15:00 (00:00-08:00) - UTC (PT)

State of H1: Lock acquisition

Intent Bit: Commissioning

Weather: 0-30 mph wind

Primary 0.03 – 0.1Hz: 0.01 um/s

Secondary 0.1 – 0.3Hz: 0.1 um/s

Outgoing Operator: Travis

Quick Summary: Coordinating with Dave and Jason to try and recover the IFO. Still seeing a good deal of lightning on the cameras.

Comments related to this report
keita.kawabe@LIGO.ORG - 02:51, Wednesday 24 July 2019 (50767)

Received a call from Niko.

EY problem -> Wait for Dave till the morning.

Corner station problems need to be worked on now. Niko called Jason and Peter King to ask how to reset PMC HV supply.

In the mean time Niko told me that he couldn't move some of the optics state to ALIGNED. That was mostly becaue of DC centering loops and SQZ ASC were outputting huge numbers, so I just disabled ASC (didn't have to do that for SQZ ASC) and cleared the history.

MC2 length output was big because the MC board boost was on when not in lock. I disabled it to prevent MC2 from tripping. Hopefully the guardian will take care of the rest of the MC board setting once PSL comes back and when Niko runs the down script. If Niko cannot lock IMC he'll call me again.

Displaying report 1-1 of 1.