Displaying report 1-1 of 1.
Reports until 08:27, Thursday 25 July 2019
H1 PSL (PSL)
peter.king@LIGO.ORG - posted 08:27, Thursday 25 July 2019 (50802)
Outage post mortem from a PSL perspective
Looking at the trend data, the NPRO switched off on July 23rd, 2019 at 10:33:46 PM local time (1247981644 GPS time) and was off for ~26 minutes.
Attached are plots of the NPRO, front end laser, and 70W amplifier outputs.  The pre-modecleaner output and high voltage power supply output
for the same time period is attached (PMC.png and PMCKepco.png).  The pre-modecleaner was not operational again for approximately another 4 hours
and 7 minutes.  The FSS hadn't really stabilise until another ~37 minutes had passed (FSS.png).  The ISS was not on until ~11.5 minutes after
the pre-modecleaner was restored (ISSLag.png).

    The PSL was not functional until July 24th, 2019 at 2:57:42 AM local time (1247997480 GPS time), some 4 hours and 24 minutes after going
down.

    Otherwise the PSL seems okay.  It might be worth checking the performance of the servos at the next opportune maintenance Tuesday.
Images attached to this report
Displaying report 1-1 of 1.