georgia.mansell@LIGO.ORG - posted 17:54, Monday 05 August 2019 - last comment - 19:24, Monday 05 August 2019(51042)
Updating CAL-CS LSC (auxiliary DOFs) filters

WP 8302

I wanted to get some calibrated spectra of the MICH/PRCL/SRCL degrees of freedom and noticed that the DRMI FOM - i.e. the plot on nuc 4 was not calibrated correctly. The first thing to fix was the outdated actuator filters in the CAL-CS model (the no longer matched the filters that were engaged for BS/PRM/SRM).

While Jenne and Cheryl were relocking today I switched the filters to what I think they will be (will confirm once we are fully locked again).

This has caused several SDF diffs in CALCS which I have already accepted in the safe snap but will need to be accepted before we go to observe.

Comments related to this report
jenne.driggers@LIGO.ORG - 19:24, Monday 05 August 2019 (51043)

Apparently they are not monitored in Observe, because once we got to NLN, there were no SDF diffs.