Displaying report 1-1 of 1.
Reports until 16:05, Wednesday 16 January 2013
LHO VE
kyle.ryan@LIGO.ORG - posted 16:05, Wednesday 16 January 2013 (5124)
Leak tested viewports on HAM2
Indicated He baseline of 4.5 x 10-10 torr*L/sec at start -> 10 second audible blast at each conflat joint and 5 second audible blast in volume between each window and corresponding guard shutter  

Note:  There is still an untested electrical feed-through flange on south side of HAM2 between the support tubes but it is not accessible 