Preliminary work:
Input/Output filters: Installed, enabled on both ISIs
Symmetrization filters: Installed, enabled on both ISIs
Damping loops: Installed, running on both ISIs
Blend Filters:
-
750mHz blend: Installed, on both ISIs
-
250mHz blend: Installed, on both ISIs
-
100mHz blend: Installed, on both ISIs
Isolation loops:
Level 1 (UGF=10Hz, simplest):
-
Installed on both ISIs
-
Tested
-
Stable
Level 2 (UGF=25Hz, Intermediate):
-
Installed on both ISIs
-
Stable
-
Ran overnight (Jan 15-16) on both HAM2 and HAM3 ISIs*
-> We will leave HAM2/3-ISIs in this known configuration for the night.
Level 3 (UGF=35Hz, Most sophisticated):
-
Installed on HAM3-ISI today
-
Tested with the lowest blend (100mHz)
-
Will be adapted, and installed on HAM2-ISI tomorrow.
UGF = Unity Gain Frequency
* Blend filters used:
-X,Y,Z: 100mHz
-RX,RY,RZ: 250mHz