Range decrease correlating with SQZ channel

The range started decreasing around 08:30 UTC without any corresponding changes in ground motion or windspeed, so I started trending OSEMs/ other channels to look for corresponding changes. I found a candidate in H1:SQZ-DCPD_RATIO_2_DB_MON on the "optimize BLRMS" scopes (found on the SQZ overview medm) and plotted the trend next to the range, which has since come back up (see attached).

Images attached to this report
Comments related to this report
jeffrey.kissel@LIGO.ORG - 09:25, Tuesday 20 August 2019 (51406)DetChar, OpsInfo, SQZ
J. Kissel, S. Dwyer

We don't think this was a squeezer problem -- the BLRMS channel you found is a ratio of the SQZ output against the OMC DCPDs -- i.e. the source DARM error signal. Thus, this BLRMS can't distinguish between the squeezer loosing dB's and any ol' noise in the IFO.

I attach a ASD of DARM comparing just before the noise kicked in -- 2019-08-20 08:00 UTC -- and at the worst of the noise an hour later -- 09:08 UTC. One can see that the noise is some broad-band problem between 80 and 400 Hz. We would expect squeezer problems to affect ~a few 100 Hz and above up to several kHz.

Also -- I attach every trend that's canned from the squeezer team from their MEDM overview for the surrounding +/- 10000 seconds (+/-3 hrs or so), and I don't see anything that correlates with the increased noise start (~08:43 UTC) or stop (~10:53 UTC).

Worth further investigation though!
Images attached to this comment
sharan.banagiri@LIGO.ORG - 07:26, Monday 26 August 2019 (51516)

I was wondering if we know any more about this? The lasso tool on the detchar page has identified pretty strong correlations for the range drop with a couple of channels, although I don't know what exactly they measure (one of them is a H0 Vac).

https://ldas-jobs.ligo-wa.caltech.edu/~detchar/summary/day/20190820/detchar/lasso/