Reports until 11:31, Thursday 22 August 2019
H1 General
patrick.thomas@LIGO.ORG - posted 11:31, Thursday 22 August 2019 - last comment - 17:26, Thursday 22 August 2019(51452)
SDF differences after initial alignment
Per Sheila's request these are attached.
Images attached to this report
Comments related to this report
sheila.dwyer@LIGO.ORG - 17:26, Thursday 22 August 2019 (51457)

Thank you Patrick.  I had asked for this hoping to find a clue about why we loose lock in the TR_CARM steps each time we first attempt after running intial alignment.  I actually don't see anything that looks like a clue here for an incorrect setting.  I actually don't see anything here that looks suspicous.  It is odd that the TMS QPDs show up in SDF, but the settings look correct for starting the acquisition sequence.