J. Kissel
There have been many times since we re-adjusted the initial alignment settings (last done Thursday Aug 15 2019; see LHO aLOG 51307), that the ALS EX and ALS EY front-end models show SDF differences at the 1e-17 - 1e-18 level. This is a result of the python scripts we use to set the initial alignment references -- in particular the transmon (TMS) green QPD offsets --
/opt/rtcds/userapps/trunk/als/h1/scripts/
setEndGreenQPDOffsets_XARM.py
setEndGreenQPDOffsets_YARM.py
not rounding before installing to any reasonable precision before writing the numbers to EPICs.
As a quick fix, I manually rounded and entered the numbers in to the respective EPICs channels. Attached are screenshots of the SDF diffs after rounding, mostly to prove to you and me that I didn't make a typo in doing so (which would botch the initial/global alignment references -- a think I *definitely* don't want to do at 4p PDT on a Friday).
The long term fix will be to edit the above scripts, but since we don't plan on updating the initial alignment references again any time soon, the quick fix should suffice until we get to editing the script.