PSL Status Report - Weekly

Famis11030  

Laser Status:
    Front End Power is 32.36W (should be around 30 W)
    70W Output Power is 70.7W
    Front End Watch is GREEN
    70W Watch is GREEN

    PMC:
    It has been locked 18 days, 23 hr 38 minutes (should be days/weeks)
    Reflected power = 10.95Watts
    Transmitted power = 53.81Watts
    PowerSum = 64.76Watts.

    FSS:
    It has been locked for 0 days 0 hr and 55 min (should be days/weeks)
    TPD[V] = 5.002V (min 0.9V)

    ISS:
    The diffracted power is around 1.7%
    Last saturation event was 0 days 0 hours and 53 minutes ago (should be days/weeks)


    Possible Issues: None