Displaying report 1-1 of 1.
Reports until 07:25, Monday 23 September 2019
H1 PSL
thomas.shaffer@LIGO.ORG - posted 07:25, Monday 23 September 2019 (52078)
PSL Status Report - Weekly

FAMIS11031

    Laser Status:
    Front End Power is 32.45W (should be around 30 W)
    70W Output Power is 69.49W
    Front End Watch is GREEN
    70W Watch is GREEN

    PMC:
    It has been locked 3 days, 14 hr 54 minutes (should be days/weeks)
    Reflected power = 11.08Watts
    Transmitted power = 53.25Watts
    PowerSum = 64.32Watts.

    FSS:
    It has been locked for 0 days 5 hr and 40 min (should be days/weeks)
    TPD[V] = 4.951V (min 0.9V)

    ISS:
    The diffracted power is around 2.3%
    Last saturation event was 0 days 5 hours and 39 minutes ago (should be days/weeks)


    Possible Issues: None

Displaying report 1-1 of 1.