【Craig, Evan】
Craig already logged here that we added two more stages of whitening on AS A RF45 and then accepted the SDF differences for the resulting antiwhitening filters that have to be engaged.
Note for operators: if you find that this change prevents lock acquisition, you can revert it by going to sitemap > ASC > ASC Overview > AS_A_RF45 (first red button up in the left-hand corner) -> Q filters, and then clicking off Stage 3 whitening and Stage 2 whitening (leave on Stage 1 whitening). You'll then have to accept SDF differences for the eight RF45 filter modules.
However, Craig and I looked at the ADC counts on these channels during previous lock acquisitions and we think there should be no ADC range issues.