cheryl.vorvick@LIGO.ORG - posted 11:26, Sunday 29 September 2019 (52185)
lockloss while switching to EQ mode
M 6.8 - 66km WSW of Constitucion, Chile  (wa originally pegged at M7.2)

Usualy Iswitch right away, while Z BLRMS is around 40.  Switched with Z BLRMS at around 300 (advice has been to wait, so maybe this was too long of a wait?), and CPSFF spiked to -2000 and broke the lock.