PSL Status Report - Weekly
FAMIS 11032

    Laser Status:
    Front End Power is 32.51W (should be around 30 W)
    70W Output Power is 69.63W
    Front End Watch is GREEN
    70W Watch is GREEN

    PMC:
    It has been locked 6 days, 2 hr 29 minutes (should be days/weeks)
    Reflected power = 11.68Watts
    Transmitted power = 52.66Watts
    PowerSum = 64.35Watts.

    FSS:
    It has been locked for 0 days 9 hr and 37 min (should be days/weeks)
    TPD[V] = 4.402V (min 0.9V)

    ISS:
    The diffracted power is around 2.2%
    Last saturation event was 0 days 9 hours and 37 minutes ago (should be days/weeks)