Displaying report 1-1 of 1.
Reports until 17:20, Wednesday 02 October 2019
LHO VE
chandra.romel@LIGO.ORG - posted 17:20, Wednesday 02 October 2019 - last comment - 19:25, Thursday 03 October 2019(52268)
XBM pumping down and YBM vented

{Bubba, Chris, Tyler, Chub, Jordan, Chandra}

Another good day of H2 decoupling work. We removed the expansion spool from BSC7 and installed the new door with no issues. Annulus pumped down quickly indicating no inner and outer o-ring leaks. We then blew down XBM volume (-37degC dew point) and proceeded to rough out volume. PT170 is currently reading 5 Torr. We will valve the QDP80 out for the night and resume pumping tomorrow morning.

Had time to vent YBM and remove expansion spool on BSC8. BSC8 flange is currently covered with C3 soft cover with a healthy purge air flow. The spool is connected to west crane and will be transferred to LVEA high bay area tomorrow morning when we're in full laser safe mode.

I moved the annulus pumping system from BSC7 to BSC8 in preparation for door installation tomorrow morning.

Images attached to this report
Comments related to this report
michael.zucker@LIGO.ORG - 03:17, Thursday 03 October 2019 (52273)

Well done! 

chandra.romel@LIGO.ORG - 19:25, Thursday 03 October 2019 (52295)

New BSC7 door is PN D1800205. The PN and SN were accidentally stamped on vacuum side of plate! (same with BSC8 plate)

Displaying report 1-1 of 1.