Reports until 10:20, Thursday 03 October 2019
LHO VE (CDS, VE)
patrick.thomas@LIGO.ORG - posted 10:20, Thursday 03 October 2019 - last comment - 13:31, Thursday 03 October 2019(52278)
Updated and restarted h0vaclx
WP 8396

Done to change PT180 from a BPG402 Inficon gauge to a BCG450 Inficon gauge. Changes to C# and Powershell scripts submitted to svn yesterday. Restart done today.

Did not go as smoothly as expected. Upon regenerating the solution the names of most of the channels in the I/O list for each of the BPG402 gauges changed to what I thought was the 1.3.0.0 version. However, the version listed for the gauges is still 1.2.1.0. I do not understand why this happened. For now I have changed the Powershell script to use the 1.3.0.0 version names. I am not submitting this change to svn.

Burtrestored to 8:00 am this morning.

DAQ restart and medm changes still needed.
Comments related to this report
patrick.thomas@LIGO.ORG - 10:35, Thursday 03 October 2019 (52279)
I found this alog: https://alog.ligo-wa.caltech.edu/aLOG/index.php?callRep=44196

Does /ligo/home/vacuum/alarm_set.bsh need to be run?
patrick.thomas@LIGO.ORG - 13:31, Thursday 03 October 2019 (52283)
Ran /ligo/home/vacuum/alarm_set.bsh