HAM6 ISI looks okay

I unlocked the ISI this morning, after fixing some poorly routed cables on the  east side of the ISI. ISI tfs look okay, all of the current tfs are the red, blue and green traces on the attached plots, and are refs 6-11. Refs 0-5 are from the previous closeout. There is a small wow at 3.5hz on the H3 tf (ref 11), but I don't think it's a problem. Maybe coming from one of the suspensions? JeffK told me all sus were damped, but maybe not?