H1 SEI
jim.warner@LIGO.ORG - posted 10:12, Friday 18 October 2019 - last comment - 16:34, Tuesday 21 April 2020(52560)
IMC and CPS differential measurements

This week Chiara, Jenne and I installed a bunch of infrastructure in SEIPROC to allow us to try to reduce low frequency differential motion between chambers by adding differential cps to the error points of different chambers (mostly outlined in Chiara's 52477, similar to work that Seb and Arnaud have done at LLO). We've tried running this a couple times and haven't gotten it to work. Just as an idiot check, this morning I looked at the transfer function between the differential signal and the ISC input to MC2. I'd like a chance to collect a quick driven transfer function to the IMC with HAM3, but I don't think I'll have time today.

First image are the magnitude and phase of the CPS differential (HAM3_SUSINF_X_INMON) to the sus M3 ISC input, looks reasonable, seems fittable. Second plot is the coherence pretty good, not as high as I'd like, but these are just the epics monitors so I don't totally trust it.

Images attached to this report
Comments related to this report
arnaud.pele@LIGO.ORG - 16:34, Tuesday 21 April 2020 (55944)

For reference, this was an implementation of ECR E1900330